48

TT 51 SERIES

www.krohne.com

09/2010 - 4000754201 - MA TT 51 R01 en

7.6 Electrical data for outputs and inputs

Non-incendive transmitter

Intrinsically safe transmitter

Output terminals 6, 7

Input terminals 1, 2, 3, 4, 5

U

i

= V

max

≤ 36 VDC

U

0

= U

0C

≤ 3.3 VDC

I

i

= I

max

-

I

0

= I

SC

≤ 1.8 mA

P

i

= P

max

-

P

0

≤ 1.5 mW

L

i

10 μH

L

0

500 mH

C

i

12.1 nF

C

0

500 μF

Output terminals 6, 7

Input terminals 1, 2, 3, 4, 5

U

i

= V

max

≤ 30 VDC

U

0

= U

0C

≤ 6.6 VDC

I

i

= I

max

≤ 100 mA

I

0

= I

SC

≤ 26.4 mA

P

i

= P

max

≤ 900 mW

P

0

≤ 46 mW

L

i

10 μH

L

0

25 mH

C

i

12.1 nF

C

0

11 μF