TECHNICAL DATA

7

41

TT 51 SERIES

www.krohne.com

09/2010 - 4000754201 - MA TT 51 R01 en

Measuring accuracy

Accuracy

RTD and thermocouple: for detailed information refer to

RTD and T/C

accuracy table

on page 49.

Resistance digital accuracy:

0...1000 Ω: max. of ±20 mΩ or ±0.02% of measured value

1000...4000 Ω: ±0.025% of measured value; max. 0.5 Ω

Resistance analogue accuracy: ±0.03% of span

Voltage digital accuracy: ±5 µV or ±0.01% of measured value

Voltage analogue accuracy: ±0.03% of span

Total accuracy = sum of digital and analogue accuracy, calculated as an RMS

(Root Mean Square) value

Temperature influence

RTD and Thermocouple: for detailed information refer to

RTD and T/C

accuracy table

on page 49.

Resistance: ±0.005% of span per °C or °F

Voltage: ±0.005% of span per °C or °F

Cold Junction Compensation (CJC)

In-head transmitter:

In-head transmitter:

In-head transmitter:

In-head transmitter:
±

0.5°C within ambient temperature -40...+85°C /

±

0.9°F within ambient temperature -40...+185°F

Rail-mount transmitter:

Rail-mount transmitter:

Rail-mount transmitter:

Rail-mount transmitter:
±

0.5°C within ambient temperature -20...+70°C /

±

0.9°F within ambient temperature -4...+158°F

Temperature influence CJC

±

0.005°C per °C / ±0.005°F per °F

Sensor wire influence

RTD and resistance, 2-wire: adjustable wire resistance compensation

RTD and resistance, 3-wire: negligible, with equal wire resistance

RTD and resistance, 4-wire: negligible

Thermocouple and voltage: negligible

Supply voltage influence

<±0.001% of span per V within specified limits

Long-term drift

Max. of ±0.01°C or ±0.01% of span per year /

Max. of ±0.02°F or ±0.01% of span per year

Operating conditions

Temperature

Temperature

Temperature

Temperature

In-head transmitter

Operating and storage temperature:

Operating and storage temperature:

Operating and storage temperature:

Operating and storage temperature:

Standard version: -40...+85°C / -40...+185°F

IS version: for detailed information refer to

Temperature data for areas with

potentially explosive atmospheres

on page 46.

Rail-mount transmitter

Storage temperature:

Storage temperature:

Storage temperature:

Storage temperature:

Standard version: -40...+85°C / -40...+185°F

Operating temperature:

Operating temperature:

Operating temperature:

Operating temperature:

Standard version: -20...+70°C / -4...+158°F

IS version: for detailed information refer to

Temperature data for areas with

potentially explosive atmospheres

on page 46.

Humidity

5...95% RH (non-condensing)

Protection category

Protection category

Protection category

Protection category

In-head transmitter

Housing: IP65

Terminals: IP00

Rail-mount transmitter

Housing: IP20

Terminals: IP00