TECHNICAL DATA

38

OPTITEMP TT 50 C/R

www.krohne.com

09/2010 - 4000636502 - MA OPTITEMP TT 50 C/R R03 en

7.7 RTD and T/C accuracy table

Accuracies in °C

Accuracies in °F

INFORMATION!

Conformance level 95% (2σ)

CJC = Cold Junction Compensation

Input type

Temp. range

Min. span

Accuracy

Temp. influence

(Dev. from ref. temp. 20°C)

[°C]

[°C]

[°C]

RTD Pt100

-200...+1000

10

±

0.2°C or ±0.1% of span

±

0.01% of span per °C

RTD Ni100

-60...+250

10

±

0.2°C or ±0.1% of span

±

0.01% of span per °C

T/C type J

-200...+1000

50

±

0.3°C or ±0.1% of span

1

±

0.01% of span per °C

T/C type K

-200...+1350

50

±

0.5°C or ±0.1% of span

1

±

0.01% of span per °C

T/C type S

-50...+1750

300

±

2.0°C or ±0.1% of span

1

±

0.01% of span per °C

T/C type B

+400...+1800

700

±

2.0°C or ±0.1% of span

1

±

0.01% of span per °C

1 CJC error is not included

Input type

Temp. range

Min. span

Accuracy

Temp. influence

(Dev. from ref. temp. 68°F)

[°F]

[°F]

[°F]

RTD Pt100

-328...+1832

50

±

0.4°F or ±0.1% of span

±

0.006% of span per °C

RTD Ni100

-76...+482

50

±

0.4°F or ±0.1% of span

±

0.006% of span per °C

T/C type J

-328...+1832

122

±

0.5°F or ±0.1% of span

1

±

0.006% of span per °C

T/C type K

-328...+2462

122

±

0.9°F or ±0.1% of span

1

±

0.006% of span per °C

T/C type S

-58...+3182

572

±

3.6°F or ±0.1% of span

1

±

0.006% of span per °C

T/C type B

+752...+3272

1292

±

3.6°F or ±0.1% of span

1

±

0.006% of span per °C

1 CJC error is not included