Technical data, 5 temperature data for potentially explosive areas, 6 electrical data for outputs and inputs – KROHNE OPTITEMP TT 50 C-R EN User Manual
Page 37

TECHNICAL DATA
7
37
OPTITEMP TT 50 C/R
www.krohne.com
09/2010 - 4000636502 - MA OPTITEMP TT 50 C/R R03 en
Formula for the maximum permissible output load of the TT 50 R:
permissible R
Load
[Ω] = (U-11)/0.023
7.5 Temperature data for potentially explosive areas
In-head transmitter (Ex-version)
7.6 Electrical data for outputs and inputs
In-head transmitter (Ex-version)
Output load diagram TT 50 R
1 R: total output load in Ω
2 U: supply voltage in VDC
Temperature class
Ambient temperature T
a
T6
-40°C ≤ T
a
≤ +50°C / -40°F ≤ T
a
≤ +122°F
T5
-40°C ≤ T
a
≤ +65°C / -40°F ≤ T
a
≤ +149°F
T4
-40°C ≤ T
a
≤ +85°C / -40°F ≤ T
a
≤ +185°F
Ouput (supply)
Input (sensor)
Max. voltage to
transmitter
U
i
= 30 VDC
Max. voltage from
transmitter
U
o
= 30 VDC
Max. current to
transmitter
I
i
= 100 mA
Max. current from
transmitter
I
o
= 25 mA
Max. power to
transmitter
P
i
= 900 mW
Max. power from
transmitter
P
o
= 190 mW
Internal inductance
L
i
= 1 mH
Max. inductance (input
loop)
L
o
= 19 mH
Internal capacitance
C
i
= 1 nF
Max. capacitance (input
loop)
C
o
= 31 nF
.book Page 37 Monday, September 13, 2010 5:07 PM