beautypg.com

Installation – KROHNE OPTIFLEX 2200 C_F Quickstart EN User Manual

Page 25

background image

INSTALLATION

2

25

OPTIFLEX 2200 C/F

www.krohne.com

08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en

• Attach a union nut to each coaxial tube (middle and end tubes)
• Attach a middle tube segment to the coaxial probe stem. Do not use tools to tighten the

assembled parts.

• Assemble the next middle rod segment (with a pair of lock washers on the attached screw)

and the top rod segment. Use two 7 mm open-ended wrenches to tighten the assembled parts
to a torque of 2…3 Nm.

• Assemble the next coaxial tube segment and the top coaxial tube segment. Do not use tools to

tighten the assembled parts. Do steps (9) thru (10) again until the end rod segment and end
coaxial tube are attached.

• Use the 2 pipe wrenches to tighten the coaxial tubes in the lock nuts.
• Use a 2.5 mm Allen wrench to attach and tighten the two HC M5×5 screws (lock screws) to the

union nut.

Figure 2-27: How to assemble the segmented coaxial probe: part 2

WARNING!
Be careful when you use the pipe wrenches. Make sure that the measuring tubes have no
deformation.

CAUTION!

Make sure that the screws are tight and the measuring tube cannot loosen.

Make sure that the fitting for the lock screw is not aligned with a hole in the coaxial tube.