INSTALLATION

22

OPTIFLEX 2200 C/F

www.krohne.com

08/2013 - 4000622103 - QS OPTIFLEX 2200 R03 en

• Attach a lock nut to each end of the rod probe segments.
• Attach a union nut to the bottom end of each rod probe segment, but not to the bottom

segment. Use two 8 mm open-end wrenches to tighten the union nut against the lock nut.

• Attach the top segment of the rod probe to the union nut below the process connection. Use

two 8 mm open-end wrenches to tighten the union nut against the lock nut on the rod probe.

• Attach the middle segment of the rod probe to the union nut on the top segment (if there are

middle segments). Use two 8 mm open-end wrenches to tighten the union nut against the lock
nut. Repeat this step for the other segments.

• Attach the bottom segment of the rod probe to the union nut on the top segment. Use two

8 mm open-end wrenches to tighten the union nut against the lock nut.

Figure 2-24: How to assemble the segmented single rod probe: part 2

WARNING!
Put a support below the probe to prevent deformation.

CAUTION!
Make sure that the nuts are tight and the rod probe cannot loosen.