Figure 1.14, Instruction sheet – In-Situ S2XP Restrictor for TROLL 9500—Installation and Calibration Instructions User Manual
Page 5
Instruction Sheet
Figure 1.14 Removing the pH/ORP sensor
5. Remove the conductivity sensor using the removal
tool. Note the sensor's port location (port 4) for
reinstallation.
6. Remove the RDO sensor by pulling the connector out
of the TROLL 9500 body. Never remove the sensor by
pulling on the cable.
7. Reinstall all sensor port plugs.*
8. Reinstall the pH/ORP sensor and conductivity sensor.
Ensure both sensors are completely inserted.
9. Check all sensor ports. Never ship the TROLL 9500
with open sensor ports. *
10. Fill the Cal Cup with a small amount of Quick Cal
solution. Install the Cal Cup onto the TROLL 9500. *
11. Replace red dust caps on the RDO probe and TROLL
9500 connectors.
* All sensor ports should be occupied by a
sensor or port plug prior to shipping.
DO NOT INSTALL THE CAL CUP
WITH ANY LIQUID IN IT IF YOU
HAVE LOST PORT PLUGS
.
Liquid will enter and damage the instrument.
YOU WILL BE CHARGED FOR
INSTRUMENT REPLACEMENT IF
YOU DAMAGE THE INSTRUMENT
.
Contact In-Situ if you have questions
regarding proper shipping.
800-446-7488
5
www.in-situ.com