9500 S2XP Restrictor Installation and Calibration

Part number 0085490, Rental R79420

Overview

The S2XP restrictor is shipped in two pieces. The
restrictor must be installed in separate steps to allow for
sensor calibration.

It is recommended you follow all steps in this instruction
sheet to properly set up your TROLL 9500 S2XP
system.

Contents

Figure 1.1 Contents of the shipping container

1

TROLL 9500 Instrument

2

S2XP restrictor, part A

3

Quick Cal solution

4

S2XP restrictor, part B

5

Sodium sulfite solution

6

Accessories kit

7

RDO sensor; TROLL Com

8

Bubbler (not shown)

Units may contain other equipment.

pH, ORP, and Conductivity Calibration

1. Detach the Cal Cup from the TROLL 9500. The Cal

Cup contains a small amount of liquid to hydrate the
pH sensor during shipment. Discard this liquid.

2. Thoroughly rinse the sensors with clean water.

3. Fill the Cal Cup to the fill line with Quick Cal solution.

Reinstall the Cal Cup on the TROLL 9500. See
Figure 1.2. Do not over-tighten.

Figure 1.2 Calibration using Quick Cal in the Cal Cup

4. In Win-Situ 4, click on Parameters in the Navigation

tree.

5. Click Quick Cal. Select pH, ORP, and Conductivity.

6. Click Start. You must hold the TROLL 9500 vertically

for proper calibration.

7. Allow the calibration to run until indicators reach

Stable. Check calibration coefficients to ensure a
proper calibration.

Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, iSitu, Pocket-Situ, RDO,
RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ Inc.

©

2013. All rights reserved.

08/2013 | Rev. 005 | 0085492