9 controller requirements and connections, Wiring overview, Stripped-and-tinned cable – In-Situ Aqua TROLL 400 Operators Manual User Manual
Page 32: Power connections, Controller requirements and connections

800-446-1500
32
www.in-situ.com
Controller Requirements and Connections
The instrument can be connected to a controller or logger for communication via the
following protocols.
l
SDI-12
l
RS485 Modbus
l
RS232 Modbus (with converter)
Wiring Overview
Refer to the diagrams on the following pages. Trim and insulate unused wires. The
shielded wire should be wired to a chassis ground or earth ground. See "Safety" on
page 5.
Stripped-and-Tinned Cable
Signal
Color
Ground/Return
Black
External Power
Red
No Connection
Brown
RS485 (-)
Green
RS485 (+)
Blue
SDI-12
White
Power Connections
The Aqua TROLL 400 requires an external 8 to 36 VDC power source. The red wire
must be connected to the positive terminal of the power source. The black wire must be
connected to the negative terminal of the power source, which is often referred to as the
system ground or return.