ELECTRICAL SAFETY TESTER

USER MANUAL

7

WARNING: To avoid personal injury, disconnect the

power cord before removing the fuse holder.

Operator’s Precaution

(1) With immense high output voltage and current of the puncture

tester, only qualified person can operate the tester in order to

avoid fatal electric shock.

(2) On-job training is required for operator to better use the tester

smoothly and safely.

(3) The operator is prohibited to dress with metal ornaments or wear

metal decoration in order to avoid electric shock.

(4) The person with cardiac or wear a pacemaker must not to operate

the tester.

(5) Be sure to wear electric protective gloves whenever operating

this instrument, in order to guard against electric shock hazards.

(6) Before turning on the power switch, make sure that the TEST

VOLTAGE dial is in the counterclockwise extreme position(Min.

voltage).

(7) Be sure to turn off the power switch each time the instrument is not

used even for a short period of time or when the operator leaves

the instrument.

(8) Before touching the leads or output terminals, be sure to turn off

the instrument

ELECTRICAL SAFETY TESTER

USER MANUAL

8

Secure Testing

Never operate the tester in the place with electric circuit device

around.

The earth lead should be well connected in accordance with

instruction. The Return Lead has to be connected to the tested object

first before linking up test probe. Do not plug the high voltage test

probe to the high voltage output terminal before doing the testing.

Also, do not touch the electric conductor of test probe and the

operator has to fully control the power on/off by using switch or

remote control, which should not be lay aside carelessly.

WARNING: During the testing, do not touch the

tested object or any other connected objects.

3-3. Environment

The normal ambient temperature range of this instrument is from 0

°

to 40

°C (32° to 104°F). To operate the instrument over this specific

temperature range may cause damage to the circuits.

Do not use the instrument in a place where strong magnetic or

electric field exists as it may disturb the measurement.

WARNING: This is a Class A product. In a domestic

environment this product may cause radio

interference in which case the user may be required

to take adequate measures.