ELECTRICAL SAFETY TESTER

USER MANUAL

15

(equipment under test).

12) Connect a red high pot probe from “High voltage output

terminal (17)” to the EUT (equipment under test).

13) Press START key once to appear STBY (Stand by) on the

display, then adjust (13) the voltage knob until the value of 5.00

is display on the panel.

14) Then, press START key to proceed testing, during testing, the

warning indicator (3) will be flashing continuously.

15) If the value displayed on the panel drops because of the load

adding test, adjust “Voltage knob (13)” to maintain the value at

5.00.

16) If the test on the EUT is approval, the PASS indicator on the

panel will be lighted up.

17) During testing, if the ARC is occurred, both the lights of ARC

and FAIL will be on, and the buzzer will blare out a warning,

now remove the high pot probe from the EUT, and press RESET

to turn off the buzzer, ARC and FAIL lights, then the instrument

will be back to the initial status.

18) If the test is over the jump-out current, the FAIL indicator on the

panel will be lighted up and the buzzer will blare out a warning.

Now remove the high pot probe from the EUT, and press

RESET to turn off the buzzer and FAIL light, then the

instrument will be back to the initial status.

ELECTRICAL SAFETY TESTER

USER MANUAL

16

5-3. Insulation Test

Take GPT-825 as example, set output to 1000V, but without

specifying the test time.

1) Connect the Withstanding/Insulation Tester to the ground.

2) Adjust the “Voltage Knob” counter-clockwise to the lowest

voltage position.

3) Set power switch to ON position.

4) Set the key (15) to IR position.

5) Set the Timer to OFF position.

6) Connect a test lead from the Return terminal (18) to the EUT

(equipment under test).

7) Connect a red high pot probe from “High voltage output

terminal (17)” to the EUT (equipment under test).

8) Press START key once to appear STBY (Stand by) on the

display, then, press START key to proceed testing. During

testing, the warning indicator (3) will be flashing continuously.

9) If the insulation resistance of the EUT is within 2000MΩ, thus

the panel will display the insulation resistance value of the EUT,

and the testing is kept going until the RESET key is pressed.

10) If the test on the EUT is disapproval, the FAIL indicator on the

panel will be lighted up and the buzzer will blare out a warning.

Now remove the high pot probe from the EUT, and press

RESET to turn off the buzzer and FAIL light, then the

instrument will be back to the initial status.