beautypg.com

Application – GW Instek GFC-8010H User Manual

Page 5

background image

FREQUENCY COUNTER

USER MANUAL

7

5. APPLICATION

5-1. Sensitivity

The role of the SENSITIVITY (or attenuator) switch in a common

measuring instrument is to protect the input circuit and prevent the meter

from going off scale.

For a counter, SENSITIVITY is still one of the large roles. Generally,

hysteresis occurs in the waveshaping circuit of the counter. In order for the

instrument to put up resistance to noise, the circuit will not work even

when the noise is lower than the hysteresis applied. The waveshaping

circuit is a Schmitt circuit and the operation of this circuit is described

below:

Fig. 2 Operation of the Schmitt circuit

FREQUENCY COUNTER

USER MANUAL

8

As refer to Fig. 2, when input voltage is at V+, the output voltage is high

(V

OH

), while input voltage is at V

-

, the output voltage is low (V

OL

). The

difference between these two voltage V

H

=(V

+

)-(V

-

) is called the hysteresis

voltage.

But if both V

+

and V

-

don’t react each other, no output will be obtained

and the Schmitt circuit will not work out with the states of (1), (2) and (3)

of Fig. 3 shown as below.

Fig. 3 States under which the schmitt circuit doesn’t work

From above description, it can be easily understood whether or not the

Schmitt circuit works is attributed to the SENSITIVITY (Attenuator) to

determine the magnitude of the input voltage.

An example of preventing erroneous counting by correctly selecting the

SENSITIVITY shown as Fig. 4 below:

(a) Correctly counting a distortion signal by selecting suitable

SENSITIVITY. However, when the input voltage is too high, a

frequency doubles the unknown frequency will be indicated.

(b) Erroneous counting occurs when high frequency noise is

superimposed on the unknown signal and the input voltage of the

Schmitt circuit is too high. However, a correct counting can be

obtained by selecting suitable SENSITIVITY.