Figure 15 dvm connection, 2 loop test calibrate, D/a trim – Detcon 700 Series Remote Sensor/Alarm Relay Module With HART Interface User Manual
Page 25
700 HART-RAM I.M.
NOTE 2: If the sensor fails the stability criteria, a “Stability Fault” will be declared and a “Fault Detected”
message will be displayed alternately on the sensor with the current reading. The HART Sensor Status will
change to reflect a Sensor Fault.
NOTE 3: If the sensor fails the clearing time criteria, a “Clearing Fault” will be declared and a “Fault
Detected” message will be displayed alternately on the sensor with the current reading. The HART Sensor
Status will change to reflect a Clearing Fault.
6.3.2.2 Loop Test Calibrate
The Calibration Screen contains a Loop Test Calibration section. This section displays the loop current
reading and allows the user to perform D/A trim and Loop Test. D/A trim is used to calibrate the 4mA and
20mA set points in the Digital/Analog converter. Loop test allows the user to set the mA output to any level
within range; this can be a good troubleshooting tool for line degradation. A DVM capable of reading
milliamps and a 100~250 Ohm resistor are required to perform D/A trim and/or Loop test.
+ 24V
100~250 Ohm Resistor
DVM +mA
GND (- 24V)
DVM GND
HART
Connector
HART-RAM Terminal Board
ALM1
ALM2
FLT
NO NC COM NO NC COM NO NC COM
J7
J2
J1
J4
IN
SENSOR
OUT
+
+
mA
A
B
mA
A
B
+
A
B
K1
K3
K2
Figure 15 DVM Connection
D/A Trim
1) Select “D/A trim” from the Loop test Calibrate section of the screen.
2) The procedure will prompt to connect a meter: (Refer to Figure 15.)
a) Disconnect the wire from the HART Connectors mA/HART position.
b) Connect one end of a 100~250Ohm resistor to the mA/HART position of the HART Connector.
c) Connect the positive lead of a DVM set to measure milliamps to the other end of the resistor.
d) Connect the negative lead (-) of the DVM to the GND position of the connector.
3) The procedure will set the output to 4mA, and prompt for the DVM Reading. Input the reading and
select “OK”.
4) The HART-RAM will readjust the output for 4mA, and inquire if the reading is 4mA on the DVM
(select “YES” or “NO”).
5) If the reading on the DVM is 4mA, select “YES” and “OK” and the HART-RAM will continue on to
20mA. If the reading is not 4mA, select “NO” and “OK” and the procedure will re-run the 4mA
calibration (Step 3).
6) The procedure will set the output to 20mA, and prompt for the DVM Reading. Input the reading and
select “OK”.
7) The HART-RAM will readjust the output for 20mA, and inquire if the reading is 20mA on the DVM
(select “YES” or “NO”).
8) If the reading on the DVM is 20mA, Select “YES” and “OK”. If the reading is not 20mA, select “NO”
and “OK” and the procedure will re-run the 20mA calibration (Step 6).
9) When both the 4mA and the 20mA have been calibrated the procedure will return the HART-RAM to
automatic control.
700 HART-RAM I.M.
Rev. 0.0
Page 21 of 24