Figure 8 hart-ram software flowchart – Detcon 700 Series Remote Sensor/Alarm Relay Module With HART Interface User Manual
Page 10
700 HART-RAM I.M.
The user interface of the HART-RAM is designed to mimic that of the Model 700 Gas Sensor. However,
only the functions deemed critical for normal remote sensor operation are available. These are the 5 menu
functions that are available for the remote control of the 700 Gas Sensor:
AutoZero – used to perform AutoZero remotely
AutoSpan – used to perform AutoSpan remotely, user is required to apply span gas flow to remote gas
sensor
View Program Status – displays the complete list of sensor status and diagnostic indicators
Set AutoSpan Level – used to change the span gas concentration
Signal Output Check – used to generate simulated outputs from the sensor for system diagnostic purposes
NOTE: For any other required operational changes, the 700 Gas Sensor must be accessed
directly.
Software Flowchart
dec
dec
Yes
/No
Set Energized
PGM1/2 (S)
PGM1/2 (3)
Yes
/No
Yes
/No
inc
Set Ascending
PGM1/2 (3)
PGM1/2 (S)
Set Latching
PGM1/2 (3)
PGM1/2 (S)
Set Level - XX
Set Alarm 2
Auto Time-Out
PGM2 (S)
PGM1/2 (3)
PGM1 (S)
PGM1/2 (M)
PGM1/2 (3)
Calibration Mode
(Auto Span)
View Sensor Status
Auto Time-Out
PGM1/2 (M)
PGM1/2 (3)
inc
PGM1 (S)
PGM1/2 (3)
Set AutoSpan Level
PGM1/2 (M)
PGM1/2 (3)
PGM2 (S)
XX
AutoTime-out
PGM2 (3)
PGM1 (3)
Normal Operation
PGM2 (3)
PGM1 (3)
Calibration Mode
(Auto Zero)
PGM1 - Program Switch Location X1
PGM2 - Program Switch Location X2
Signal Output Check
Auto Time-Out
(S) - Momentary Swipe
(M) - Momentary hold of Magnet during text
scroll until arrow prompt appears, then
release
(3) - 3 second hold from arrow prompt
(10) - 10 second hold from arrow prompt
Auto Time-out - 5 seconds
inc - Increase
dec - Decrease
X, XX, XXX - numeric values
Set Energized
Auto Time-Out
Yes
/No
PGM1/2 (3)
Yes
/No
PGM1/2 (3)
PGM1/2 (S)
PGM1/2 (M)
Set Latching
PGM1/2 (S)
PGM1/2 (3)
Set Fault
LEGEND:
Simulation
PGM1/2 (3)
PGM1/2 (M)
PGM2 (10)
Set Energized
Set Latching
Set Ascending
Auto Time-Out
Set Level - XX
Auto Time-Out
PGM1 (S)
PGM1/2 (3)
inc
PGM1/2 (3)
PGM2 (S)
XX
Set Serial ID
PGM1/2 (M)
dec
Yes
/No
Yes
/No
Yes
/No
inc
PGM1/2 (3)
PGM1/2 (3)
PGM1/2 (S)
PGM1/2 (3)
PGM1/2 (3)
PGM1/2 (S)
PGM1/2 (S)
PGM1/2 (3)
PGM2 (S)
PGM1 (S)
dec
Set Alarm 1
PGM1/2 (M)
Voltage = X.XX V
Temp = XX C
mA Output = XX.XX
Sensor Diags
Serial ID
Sensor Life XXX%
Last Cal XX Days
Range XXX ppm
AutoSpan @ XX
Version X.XX
Model Type
Alarm 2 Settings
Alarm 1 Settings
Fault Settings
Figure 8 HART-RAM Software Flowchart
700 HART-RAM I.M.
Rev. 0.0
Page 6 of 24