Detcon 700 Series Remote Sensor/Alarm Relay Module With HART Interface User Manual
Page 24
700 HART-RAM I.M.
Auto Span consists of entering “Autospan Cal” and following the display. The procedure will ask for the
application of span gas. The applied gas concentration must be equal to the Autospan gas level setting. The
factory default setting and recommendation for span gas concentration is normally 50% of range. If a span
gas containing the recommended concentration is not available, other concentrations may be used as long as
they fall between 5% and 100% of range. However, any alternate span gas concentration value must be set
in the “Auto Span Level” field before proceeding with “Autospan cal”.
CAUTION: Verification that the calibration gas level setting matches the calibration span gas
concentration is required before executing “Autospan Cal”. These two numbers must be equal.
Refer to the appropriate sensor manual for more information.
1. If applicable install the Calibration Adapter or Splash Guard Adapter with integral Cal Port.
2. Verify that the Auto Span Level is equal to the calibration span gas concentration. If the Auto Span
Level is not equal to the Calibration span gas concentration, adjust the Auto Span Level.
3. Connect the Cal Gas to the sensor, but do not apply the gas.
4. Select “Autospan Cal” from the Sensor Calibration section of the screen.
NOTE: Upon entering calibration the 4-20mA signal drops to 2mA and is held at this level until the
program returns to normal operation.
5. Upon entering the procedure the procedure will prompt to begin Auto Span Calibration.
6. The procedure will prompt to apply span gas. Apply span gas from the attached cal gas cylinder and
respond to the prompt.
NOTE: The sensor reading will respond to the gas and will switch to displaying a flashing “XX”.
NOTE: Assuming acceptable sensor signal change, after 1 minute the reading will auto-adjust to the
programmed Auto Span level. During the next 30 seconds, the Auto Span sequence checks the sensor for
acceptable reading stability. If the sensor fails the stability check, the reading is re-adjusted back to the Auto
Span level and the cycle repeats until the stability check is passed. Up to three additional 30-second stability
check periods are allowed before the unit reports a “Stability Fault” twice and the sensor will return to
normal operation, aborting the Auto Span sequence. The sensor will continue to report a “Stability Fault”
and will not clear the fault until a successful Auto Span is completed.
NOTE: If the sensor passes the stability check, the sensor reports a series of messages:
“Span OK”
“Sensor Life XXX%”
“Remove Span Gas”
NOTE: When calibrating O2 deficiency sensors, there is no requirement to clear to <5% of range. The
sensor will return to normal operation immediately after span adjustment.
7. When the sensor passes calibration the procedure will prompt to remove the span gas. Unsuccessful
completion of the span calibration will create a Global Fault, and “Autospan Cal” will be aborted with a
change to the HART Sensor Status (refer to section 6.2.2 Sensor Status).
8. After successfully setting span cal the sensor and the HART Interface will return to Automatic Mode.
9. The Auto Span Calibration is complete
10. Remove the cal gas and calibration adapter if applicable.
NOTE 1: If the sensor fails the minimum signal change criteria, a “Range Fault” will be declared and a
“Fault Detected” message will be displayed alternately on the sensor with the current reading. The HART
Sensor Status will change to reflect a Range Fault.
700 HART-RAM I.M.
Rev. 0.0
Page 20 of 24