DeFelsko PosiTector 6000 v.7.3 Full Manual User Manual

Page 5

background image

2

If memory is ON while continuous measurements are being taken,
only the last value on the display (when the probe is lifted) is
stored into memory. Scan (pg.15) stores ALL measurements into
memory.

Certification

All probes include a Certificate of Calibration. For organizations
with re-certification requirements, instruments may be returned at
regular intervals for calibration. DeFelsko recommends that
customers establish calibration intervals based upon their own
experience and work environment. Based on our product
knowledge, data and customer feedback, a one year calibration
interval from either the date of calibration, date of purchase, or
date of receipt is a typical starting point. Written calibration
procedures are available from DeFelsko Corporation at
no charge.

Golden Rule

Measure your uncoated part first! This quick zero-check
determines if a Calibration Adjustment is needed for your
substrate. (see pg. 10)

Next, lay the included plastic shims onto a bare surface and
measure them individually to ensure the Gage measures a known
thickness within tolerance.

Protective Cap

Some PosiTector 6000 models are shipped with a protective
plastic cap over the probe. Remove this cap prior to use.
Replace it when the Gage is not in use to protect the probe.

Wrist Strap

We recommend attaching and wearing the
supplied wrist strap.

1

2

Plastic Lens Shield

The LCD is covered with a thin plastic film for protection against
fingerprints and other marks during shipment. This film, while
usually removed before using the Gage, can be left in place to
protect against paint overspray. Replacements can be purchased.