beautypg.com

DeFelsko PosiTector 6000 v.7.3 Full Manual User Manual

Page 15

background image

12

When selected, the icon will appear and all calibration settings
are “locked” to prevent further user adjustments. Uncheck to make
further adjustments.

Cal Lock

ü

Select N Lock (Non-Ferrous Lock) when operating regularly on
non-ferrous substrates. The icon will appear and the probe will
only use the eddy current principle to shorten measurement time
and extend battery life.

N Lock is useful when measuring coatings on plated steel.
Normally the probe measures both the coating + plating over the
steel using the magnetic principle. N Lock makes the Gage
measure the coating over the plating only.

Select N Lock to measure over slightly magnetic substrates; i.e.
clear-coat on gold over nickel-plated brass. Although the probe's
magnet is used for the magnetic principle, it is also used in N Lock
to magnetically saturate a slightly magnetic substrate and allow
the eddy current principle to operate unhindered.

With N Lock engaged it is possible to obtain a reading

when measuring non-conductive coatings on steel. This is not
recommended.

N Lock

ü

(FN ferrous/non-ferrous combination probes only)

CAUTION:

4. Lift the probe from the surface and adjust the displayed reading
down (-) or up (+) to the known thickness value of the thin item. Press
to accept this value.

5. Repeat steps 2 - 4 for the thicker item.