beautypg.com

14 setup menu setup menu – DeFelsko PosiTector 6000 v.7.3 Full Manual User Manual

Page 17

background image

14

Setup Menu

Setup Menu

Reset

Reset (soft reset) restores factory settings and returns the Gage
to a known condition. The following occurs:

- All batches, stored measurements, images, and batch names

are erased.

- The icon will appear on the display.

- Menu settings are returned to the following:

Memory = OFF
High Res = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF

Cal Lock = OFF
Bluetooth = OFF
Display = None
N Lock = OFF

Perform a more thorough Hard Reset by powering down the
Gage, waiting several seconds, then simultaneously holding both
the center and (+) buttons until the Reset symbol
appears. This returns the Gage to a known “out-of-the-box”
condition. It performs the same function as a menu Reset with
addition of:

NOTES:

- Keep the Gage away from metal during a Reset.
- Date, Time and WiFi are not affected by either Reset.

Units = microns
Flip Display = Normal
Auto Sync = OFF
White on Black = OFF
Fast Mode = OFF

Language = English
Battery Type = Alkaline
Backlight = Normal
Bluetooth Streaming = OFF
USB Drive= ON

- All calibration adjustments and Cal Memory are cleared and

returned to the Gage’s factory calibration settings.

- Bluetooth Pairing info is cleared.

- Menu settings are returned to the following: