LumaSense Technologies INNOVA 1303 User Manual
Page 54

Chapter 4
________________________________________________________________________
_______________________________________________________________________
BE1085-14
1303 Multipoint Sampler and Doser
LumaSense Technologies A/S
Page 54 of 60
Table 4.3 The interface jobs which control the 1303’s doser system
Job Header
Data
Effect on 1303
Calibrate_Nozzle
Specific dosing-
nozzle number.
Starts the auto-
matic calibration of
the specified doser
nozzle.
Main_Dosing_Valve
OPen
CLose
Opens/closes the
main dosing valve.
Open_Dosing_Valve
Up to 6 dosing-
valve numbers.
Opens specified
dosing valves.
Discontinuous_Dosing
Dosing valve
number, total dos-
ing time, discon-
tinuous dosing pe-
riod-length, dosing
valve opening pe-
riod.
Starts a discontin-
uous dosing proce-
dure through the
specified dosing
valve.
Dosing_Gas_Pressure?
No data.
Measures and
reads-out the trac-
er-gas pressure
(kPa) in the doser
manifold.
Dosing_Gas_Temperature? No data.
Measures and
reads-out the trac-
er-gas temperature
(°C) in the doser
manifold.
Dosage_Given?
Specific dosing-
valve number. If
no number is giv-
en, the total dos-
age through all
valves is read-out.
Reads-out the dos-
age of tracer-gas
(mg) delivered
through the speci-
fied doser nozzle
since the last read-
out of the dosage
delivered.
Dosing_Pump
On
Off
Auto
Starts the dosing
pump.
Stops the dosing
pump.
See end of
section
3.6.2
.
4.2.4
Temperature Measurement Jobs
This job controls temperature measurement with the 1303.