beautypg.com

6 warning level / switch-off level, 7 emissivity slope k: autofind – LumaSense Technologies ISR 6 Advanced User Manual

Page 32

background image

ISR 6 Advanced Manual

Software InfraWin · 32

5.6 Warning Level / Switch-Off Level

In addition to the relative signal bar graph, the warning level and the switch-off level are
indicated as vertical lines in colors corresponding to the windows below.

The ISR 6 pyrometers are equipped with a warning level “dirty window” monitoring system.

A correct temperature measurement might be impossible if the ratio pyrometer is working at a
too low signal level. To avoid these wrong measurements in advance, a warning signal can be
set to a certain contamination level. A built-in relay switch can be used to switch to a warning
signal when the signal becomes too low. When this happens, the built-in indicator of the
pyrometer starts blinking, but will continue the measurement.
The warning level can be set between 0 and 99%. 0% means the “dirty window” warning
system is switched off (factory setting) and the relay can perform the the function external
clearing of maximum value storage or external hold, when one of them is activated.

The switch-off level is a function used to avoid measuring errors caused by too low signals.

Ratio pyrometers are able to measure temperatures correctly even with very low signals. For
example, through a dirty viewing window, or if dust exists in the field of view, or if the spot is
not filled by the measuring object.
If the signal is too low for a correct measurement, the pyrometer interrupts the measurement
and displays 1° below the beginning of the temperature range. Although the factory default is
set at 10%, the switch-off limit can be adjusted between 2% and 50%, depending on the
application.

5.7 Emissivity Slope K: AutoFind

InfraWin also offers an input field K for the emissivity slope. If the emissivity is changed, the
temperature change connected with this can be directly affected.

If the true temperature of the measured object is known, you can calculate the emissivity slope

of the measured object using the K: AutoFind function:

A measured temperature is displayed with
the current set emissivity slope. In this
example, the emissivity slope is set at 1.000
and the resulting temperature reading is
845.2 °C.
If you press the K: Autofind button, a

window will open which allows you to
enter the true temperature.
In this example, we have entered a
temperature of 870 °C.
Once you enter the temperature value, you
can press the OK button to set the

temperature value and close the
temperature window.