Power Consumption:
Max. 3 W ( incl. laser)
Load (analog output):
0 to 500 W
Isolation:
Power supply, analog output and digital interface are electrically
isolated from each other
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see
or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4 Dimensions
2.5 Physical User Interface
ISR 6 Advanced
with Through Lens
ISR 6 Advanced
with Laser
1 12-Pin Connector
2 Digital Display
3 Sighting Option
4 Focus Adjustment
Set Screw
All dimensions in mm
5 LED Distance
Indicator Light
6 LED 1-Channel
Mode Indicator
Light
7 Viewfinder
Adjustment Ring