Optics
Sighting:
Built-in laser aiming light (max. power level < 1 mW,
l = 630-680 nm, CDRH class II) or through-lens sighting
Optics:
Manually focusable from rear cover with
measuring distance a = 210 to 5000 mm
Distance ratio:
MB 14
approximately 100 : 1
MB
18
approximately
190
:
1
MB 25 and MB 30
approximately 350 : 1
Environment
Protection Class:
IP 65 IEC 60529 (value in mated condition)
Operating Position:
Any
Ambient Temperature:
Note: During operation the
instruments will warm up and might
reach an intrinsic temperature of up
to 58 °C
0 to 65°C at housing
Storage Temperature:
-20 to + 80°C
Relative Humidity:
Non condensing conditions
Weight:
0.6 kg
Housing:
Stainless steel
CE-label:
According to EU directives about electromagnetic immunity
Interface
Connection:
12-pin connector
Display (in rear cover):
LED, 4 digit matrix, 5 mm high
for 2-color or 1-color temperature signal or measuring distance
Parameters:
Adjustable via interface: 2-color / 1-color temperature signal,
accordingly emissivity slope or emissivity, temperature sub range,
settings for maximum value storage, address, baud rate, switch off limit,
warning level lens contamination monitoring system, transmittance,
response time t
90
, 0 to 20 mA or
4 or 20 mA analog output range, °C/°F.
Readable via interface: measured value, internal temperature of the
unit, measuring distance.
Communication
Analog Output:
Adjustable 0 to 20 mA; or 4 to 20 mA, linear
Digital Interface:
RS485 addressable (half-duplex)
Baud rate: 1200 to 115.2 kBd
(on request RS232, not addressable)
Switch Off Limit:
2% to 50% (adjustable via interface)
“Dirty Window” Warning:
Relay contact, max. continuous current 0.4 A, setting of the warning
level: 0 (off) to 99%
Exposure Time t
90
:
<2 ms (with dynamical adaption at low signal levels); adjustable to min;
0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
Maximum Value Storage:
Built-in single or double storage. Clearing with adjusted time t
clear
(off; 0.01
s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), via interface, automatically with the next
measuring object, external contact, hold-function
Electrical
Power Supply:
24 V DC ± 25%, ripple must be less than 50mV