<3. Parameters>
3-19
IM 11M12D01-51E
HART Communicator
Display
Ctrl card
Control card abnormality
Ave-a oxygen
Ave-a oxygen concentration outside alarm limits
Ave-b oxygen
Ave-b oxygen concentration outside alarm limits
Ave-c oxygen
Ave-c oxygen concentration outside alarm limits
Process gas
Process gas alarm contact signal received by AV550G
Cal gas press lo Cal gas press lo alarm contact signal received by AV550G
Inside temp
AV550G case internal temperature over alarm limit
Cell voltage
Cell emf abnormal
Heater temp
Heater temperature abnormal
Ch card
Channel card abnormality detected
Card comm.
Communications between control card and channel card
abnormal
Oxygen
Channel oxygen concentration outside alarm limits
Zero conc. ratio
Zero correction factor outside normal range
Span conc. ratio Span correction factor outside normal range
Cal time over
Cell emf didn't stabilize during calibration interval
CJ temp
Cold (reference) junction temp. outside normal range
Asymmetry alarm Abnormal change in asymmetry correction factor
Explanation
AV550G
display
Error 4
Alarm 1
Alarm 6
Alarm 7
Alarm 9
Error 1
Error 2
Error 3
Error 5
Alarm 1
Alarm 2
Alarm 3
Alarm 4
Alarm 5
Alarm 8
Table3.1 AV550G Alarms and Errors displayed on HART Communicator
T0301.ai
Control card
Channel card