2 setting parameters, Setting parameters, Burst mode – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 20
<3. Parameters>
3-7
IM 11M12D01-51E
3.2 Setting Parameters
3.2.1 Burst Mode
The AV550G continuously sends the data stored in it when burst mode is set “On”. Either one of
measured value, % output value, or 4 to 20 mA output value can be selected and sent. The data
is sent periodically at 75 ms intervals as a digital signal when the AV550G is set in burst mode.
Therefore, communication by the HART simultaneous communicator is also possible.
Setting of Burst Mode
Call up “
Burst option” display.
1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output --->
4.Burst option
Display
Operation
(ENTER)
Set data to be sent.
Measured value (PV)
% output value
(% range/current)
4 to 20 mA output value
(Process vars /crnt)
Call up "
Burst mode" display.
3.HART output 3.Burst mode
1
(ENTER)
Set
"
On
"
and press ENTER
(F4).
2
(SEND)
Press SEND (F2).
F4
F4
F2
F030209-1.ai
AV550G:
Burst option
PV
% range/current
Process vars/crnt
ENTER
ESC
HELP
AV550G:
Burst mode
Off
ENTER
ESC
off
on
AV550G:
HART output
1 Poll addr 0
2 Num req preams 5
3 Burst mode
On
4 Burst option
PV
HOME
SEND
HELP
********