Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual
Page 22
![background image](/manuals/360137/22/background.png)
<3. Parameters>
3-9
IM 11M12D01-51E
1. If “Polling” is set to “Never Poll” after the address is set, “Online Menu” cannot
be called up and displayed. Be sure to set “Polling” to “Digital Poll” after setting
the polling address.
2. When the same polling address is set for two or more field devices in multidrop
mode, communication with those field devices is disabled.
Example: Communication when set in the multidrop mode.
4
Select "
Configure
Communication "
.
5
Select "
Polling " .
6
x3
(ENTER)
Select "
Digital Poll"
and press ENTER (F4).
F4
F030212-1.ai
HART Communicator
Utility
1 Configure Communic
2 System Information
3 Listen for PC
4 Storage Location
5 Simulation
HART Communicator
Configure Communica
1 Polling
2 Contrast
3 Off Time
4 Ignore diagnostics
5 Delete Configs
HELP
HART Communicator
Polling
Ask Before Polling
Never Poll
Ask Before Polling
Always poll
Digital poll
ENTER
ESC
HELP
NOTE
1
2
3
(1) The HART Communicator searches for field
devices is set in multidrop mode when the HART
Communicator is turned on.
When a HART Communicator is connected to a
field device, its tag will be displayed (display 1).
(2) Select desired field device. After that, normal
communication with the selected field device is
possible. However, the communication speed is slow
in this case (display 2).
(3) To communicate with another field device, call up
display 3 , and select "Online".
(4) Display1 will appear. Repeat the above operation.
F030213-1.ai
HART Communicator
Online
1 1: ZIA-01A
2 2: ZIA-02A
rm
AV550G:
1
Device setup
Online
2 PV
3 PV URV
4 PV LRV
21.64
21
0
%
%
%
SAVE
ZIA-01A
BACK
Display
Operation
HART Communicator
1 Offline
2 Online
3 Frequency Device
4 Utility