<3. Parameters>
3-8
IM 11M12D01-51E
3.2.2 Multidrop Mode
Field devices in multidrop mode refer to the connection of several field devices on a single
communication line. Up to 15 field devices can be connected when set in multidrop mode. To activate
multidrop communication, each field device address must be changed to a unique number in the
range 1 to 15. This change deactivates the 4 to 20mA output and changes it to 4mA.
Setting of Multidrop Mode
Call up “
Poll addr” display.
1.Device setup ---> 4.Detailed setup ---> 4.Output condition ---> 1.HART output
--->1.Poll addr
Call up
“Auto Poll” display.
HART output ---> Online Menu ---> Main Menu ---> 4.Utility ---> 1.Configure Communication
---> 1.Polling
"
1
"
(ENTER) (SEND)
Set the polling address (a
number from 1 to 15) and
press
ENTER (F4).
Then Press
SEND (F2).
F4
F2
F030210-1.ai
AV550G:
Poll addr
0
0
ESC
DEL
HELP
ENTER
Display
Operation
1
(HOME)
Return "
Online Menu" with
HOME (F3).
2
Return to "
Main Menu " with
a "
previous" key.
3
Select "
Utility" .
F3
F030211-1.ai
JLK
4
AV550G:
HART output
1 Poll addr
1
2 Num req preams 5
3 Burst mode
On
4 Burst option
PV
HOME
SEND
HELP
AV550G:
1
Device setup
Online
2 PV
3 PV URV
4 PV LRV
21.64
21
0
%
%
%
SAVE
HART Communicator
1 Offline
2 Online
3 Frequency Device
4 Utility
Display
Operation