Parameters lists, Parameters lists -1 – Yokogawa Wireless Temperature Transmitter YTA510 User Manual
Page 36

4. PARAMETERS LISTS
IM 01C50T01-01E
4-1
4.
PARAMETERS LISTS
Initial setting value marked with “ * ” is as of specified upon ordering.
Item
Selection/ Setting Range
Initial Setting
HART
Communicator
Up to 8 characters
Up to 8 characters
Up to 16 characters
Up to 32 characters
mm/dd/yy
0 to 16777215
°
C, K,
°
F, or
°
R (see note 1)
Set the calibration range using the keypad.
Value for 4 and 20 mA signal is set with actual input
applied.
0 to 99 seconds.
0 to 99%.
“Sensor1”, “Terminal Temperature(term)” or “Sensor1-
Term”.
For YTA320, above plus “Sensor2”, “Sensor2-Term”,
“Average” or “Differential” temperature.
Sensor1-Sensor2 or Sensor2-Sensor1.
Specify input type.
2, 3 or 4 wire.
°
C, K,
°
F, or
°
R. (See note 1).
Enable or disable
°
C, K,
°
F, or
°
R. (See note 1).
Internal CJC, Constant CJC
-10000 to 10000
Specify input type.
2, 3 or 4 wire.
°
C, K,
°
F, or
°
R. (See note 1).
Enable or disable
Enables/disables sensor backup mode.
Shows the backup status. (Waiting or Start.)
Enables the recovery to sensor1 during backup
operation.
Tag
Tag Ex
Descriptor
Message
Date
Snsr1(2)
snsr s/n
PV unit
(SV,TV,4V)
LRV/URV
Apply values
PV Damp
(SV,TV,4V)
PV Damp point
PV is
(SV,TV,4V)
Diff direction
(YTA320 only)
Snsr1 Type
Snsr1 Wire
Snsr1 unit
Snsr1 Match
Enbl (Note 3)
Term Unit
CJC Select
CNST CJC
Temp
Snsr2 Type
Snsr2 Wire
Snsr2 unit
Snsr2 Match
Enbl (Note 2)
Snsr Bkup
Bkup state&Cntl
Bkup Return
Snsr1
Memory
Process
Variable
Sensor1
Configuration
Terminal
Temperature
Sensor2
Configuration
(YTA320 only)
Engineering Unit
Range
Damping time
constant
Damping holding
point
Variable
mapping
Differential
Direciton
Sensor1 type
Sensor1 wire
Sensor1 unit
Enable sensor
match function
Terminal Temp.
unit
CJC Function
Sensor2 type
Sensor2 wire
Sensor2 unit
Enable sensor
match function
Sensor Backup
Function
T0401_1.EPS
Note 1:
°
F and
°
R are available only when option code /D2 is specified.
Note 2:
Appears only when Sensor Matching Function(/CM1) is specified.
—*
—
—
—
—
—
°
C*
LRV:0*
URV:100*
—
2 s
2%
PV:
Sensor1
SV(YTA320):
Sensor2
Sensor1-Sensor2
Pt100*
3 wire*
°
C*
disable
°
C*
Internal CJC
0
Pt100*
3 wire*
°
C*
disable
disable
—
—