<6. ELECTRONIC TRANSMITTER (-E)>

6-16

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

6.2.10 Revision indication (F61/F62)

Functions F61 and F62 enable the indication of revision states for hardware, software of calibration EEPROM
and internal EEPROM.
The indication is called up as follows:

Description

Selection

Key

Indication

Change to setting mode

Enter

Display mode
F1-

Setting mode
Revision

5 x →

Enter

2 x →

F6-
F61
Hhh

1

Fff

2

Setting mode
EEPROM Revision

Enter

F61
F62
Aaa

3

Ccc

4

Back to display mode

F6-
Display mode

1

H = Hardware

2

F = Firmware

3

A = Internal EEPROM

4

C = Calibration-EEPROM

6.2.11 Current output test (F63)

Function F63 sets the output current to 0/4mA or 20mA respectively.
With this you can determine whether output current correction by function F32 is required.
The adjustment of the current output is as follows:

Description

Selection

Key

Indication

Change to setting mode

Enter

Display mode
F1-

Setting mode

5 x →

Enter
Enter

F6-
F61
F63

Selection
or

Output 0/4 mA
Output 20 mA

Enter

Enter

F63 0/4
F63 20
F63

Back to display mode

F6-
Display mode

Note:

During selection you can switch between 0/4mA and 20mA with the"→"-key.

If you press “

” instead of “Enter”, you can return from the selected point to the previous

menu item without activating the displayed parameter.