beautypg.com

2 installation, 3 marking, 2 installation -9 – Yokogawa RAMC User Manual

Page 101: 3 marking -9, Hazardous area safe area, Name plates of electronic transmitter

background image

<10. EXPLOSION-PROTECTED TYPE INSTRUMENTS>

10-9

IM 01R01B02-00E-E 10th edition January 01, 2013 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

Rota Yokogawa

D-79664 Wehr

WT-MAG Mat. No. 16-8040

Serial No, xxxxxxxx

Ex ia IIC T6 Gb

IECEx PTB 12.0020

Ui=30V Ii=101mA Pi=1.4W

Li=0.15mH Ci=4.16nF

10.5.2 Installation

Fig. 10.2

Ex-Version according IECEx (Option /ES1) with electronic transmitter, power supply and

limit switches with transmitter relay

15

4
6

3

1

14

12

10

11

8

7

9

12

11

10

9

I -

I +

L/+

N/-

O -

O+

Output 4-20mA

U

F

G

RAMC

A

Supply 230V AC

KFA6-SR2-Ex2.W

Option: /W2B two channels: KFA6-SR2-Ex2.W

Transmitter relay

Limit MIN

Limit MAX

Option / Kn

Rotameter

4-20mA

Max

Min

Limit switch

EN 60947-5-6 (Namur)

One channel connection like limit MAX

I = 101m A

P = 1.4W

C = 4.16nF

U = 30V

L = 0.15mH

Ex ia IIC T6 Gb

IECEx PTB 12.0020

Tamax = 70°C

Tamax = Temperature in indicator

C = 150nF

EEx ia IIC T6

L = 0.15uH

SC 3.5-NO

P = 64mW

I = 25mA

U = 16V

Temperature

Class

Transmitter power supply

Option: /W2A one channel: KF

A6-SR2-Ex1.W

KFA6-SR2-Ex2.W

IECEx PTB 11.0031X

IECEx PTB 11.0091X

Transmitter WT-MAG

T4 Tamax = 89°C

T3 Tamax = 89°C

T2 Tamax = 89°C

T1 Tamax = 89°C

T5 Tamax = 61°C

T6 Tamax = 49°C

-

+

-

+

-

+

~

~

+

-

+

-

i

i

i

i

i

i

i

i

i

F6.EPS

i

C = 2320nF

I = 19.1mA

P = 51mW

U = 10.6V

L = 97mH

[Ex ia]) IIC

o

o

o

o

o

Hazardous area

Safe area

e.g. Standard Type 2

10.5.3 Marking

Name plates of electronic transmitter :