General specifications, 1 standard specifications, General specifications -1 – Yokogawa EJA310A User Manual
Page 60: 1 standard specifications -1
IM 01C21D01-01E
10-1
10. GENERAL SPECIFICATIONS
10. GENERAL SPECIFICATIONS
10.1
Standard Specifications
Refer to IM 01C22T02-01E for F
OUNDATION
Fieldbus communication type and IM 01C22T03-
00E for PROFIBUS PA communication type
marked with “
᭛”.
᭹ Performance Specifications
See General Specifications sheet, GS 01C21D01-
00E, GS 01C21E01-00E, and GS 01C21E02-00E.
᭹ Functional Specifications
Span & Range Limits
EJA310A*:
L
M
Measurement
Span
and Range
Span
Range
Span
Range
psi
(/D1)
2.67 to
40 inH
2
O
0 to
40 inH
2
O
0.38 to
38 inHg
0 to
38 inHg
mbar
(/D3)
6.7 to 100
mmHg
(/D4)
5 to 75
0 to 100
13 to 1300
0 to 1300
0 to 75
9.6 to 960
0 to 960
MPa
0.67 to
10 kPa
0 to
10 kPa
1.3 to
130 kPa
0 to
130 kPa
A
Span
Range
4.3 to 430
0 to 430
0.3 to
30 bar
0 to 30 bar
0.3 to
30 kgf/cm
2
0 to
30 kgf/cm
2
0.03 to 3
0 to 3
T1001.EPS
* The above units are in absolute terms.
EJA430A:
A
Measurement
Span
and Range
Span
Range
psi
(/D1)
4.3 to 430
–15 to 430
bar
(/D3)
0.3 to 30
kgf/cm
2
(/D4)
0.3 to 30
–1 to 30
–1 to 30
MPa
0.03 to 3
–0.1 to 3
T1002.EPS
B
Span
Range
20 to 2000
–15 to 2000
1.4 to 140
1.4 to 140
–1 to 140
–1 to 140
0.14 to 14
–0.1 to 14
EJA440A:
C
Measurement
Span
and Range
Span
Range
psi
(/D1)
720 to 4500
–15 to 4500
bar
(/D3)
50 to 320
kgf/cm
2
(/D4)
50 to 320
–1 to 320
–1 to 320
MPa
5 to 32
–0.1 to 32
T1003.EPS
D
Span
Range
720 to 7200
–15 to 7200
50 to 500
50 to 500
–1 to 500
–1 to 500
5 to 50
–0.1 to 50
Zero Adjustment Limits:
Zero can be fully elevated or suppressed, within
the Lower and Upper Range Limits of the capsule.
External Zero Adjustment “
᭛”:
External zero is continuously adjustable with
0.01% incremental resolution of span. Span may
be adjusted locally using the digital indicator with
range switch.
Output “
᭛”:
Two wire 4 to 20 mA DC output with digital
communications, linear or square root program-
mable. BRAIN or HART FSK protocol are superim-
posed on the 4 to 20 mA signal.
Failure Alarm:
Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less
-2.5%, 3.6 mA DC or less(Optional
code /F1)
Note: Applicable for Output signal code D and E
Damping Time Constant (1st order):
The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.
Time Constant (approx. sec)
0.2
L, M, A, B, C, and D
T1004.EPS
Capsule (Silicone Oil)
Ambient Temperature Limits:
* Safety approval codes may affect limits.
–40 to 85°C (–40 to 185°F),
–30 to 80°C (–22 to 176°F) with LCD Display
Process Temperature Limits:
* Safety approval codes may affect limits.
Capsule M, A, B, C and D
–40 to 120°C (–40 to 248°F)
Capsule L for EJA310A
–40 to 100°C (–40 to 212°F)
Maximum Overpressure:
See General Specifications sheet.
Working Pressure Limits (Silicone Oil)
Maximum Pressure Limit:
See ‘Model and Suffix Codes.’
Minimum Pressure Limit:
Capsule
Pressure
L, M, A (EJA310A)
See Figure 1.
A, B (EJA430A)
See GS 01C21E01-00E
C, D (EJA440A)
See GS 01C21E02-00E
᭹ Installation
Supply & Load Requirements “
᭛”:
* Safety approvals can affect electrical requirements.
See Section 6.6, ‘Power Supply Voltage and Load