9. MAINTENANCE

Connect BRAIN TERMINAL and check self-diagnostics.

Contact Yokogawa service personnel.

F0909.EPS

Are valves opened or

closed correctly?

Is impulse piping

connected correctly?

Refer to error message summary in
Subsection 8.5.2 to take actions.

Refer to Section 6.6 for rated voltage
and load resistance.

Provide lagging and/or cooling, or allow
adequate ventilation.

Refer to Section 9.2 when selecting
instruments for calibration.

Fully open tap and stop valves, and
fully close drain valve.

Refer to individual model user manuals
and connect piping as appropriate for
the measurement purpose.

Adjust the output.

Is transmitter

installed where there is

marked variation in

temperature?

Is output adjusted correctly?

Large output error.

Does the self-

diagnostic indicate problem

location?

Are power supply

voltage and load resistance

correct?

Were appropriate

instruments used for

calibration?

YES

NO

YES

NO

NO

YES

NO

YES

NO

YES

NO

YES

NO

YES