<5. Parameter Summary>
5-5
IM 01C25T03-01E
No. Parameter name
Item
*
1
R/W
Content
Default value
Applicable
model
U
pl
oa
d
da
ta
F
P
L
M22 GASKET MATL
Gasket material
W
—
M23 PRO CON MATL
Flange material
W
—
M24 D-VENT MATL
Vent plug material
W
—
M25 PRO CON TYPE
Process connection type
W
—
M26 RS ISOL MATL
Remote seal material
W
—
M27 PRO CON SIZE
Flange size
W
—
M28 NUM RS
Number of remote seal
W
—
M29 RS FILL FLUID
Fill fluid of remote seal
W
—
M30 RS TYPE
Remote seal type
W
—
M50 MS CODE 1
Model and suffix code 1
W
—
M51 MS CODE 2
Model and suffix code 2
W
—
M52 MS CODE 3
Model and suffix code 3
W
—
M53 MS CODE 4
Model and suffix code 4
W
—
M54 MS CODE 5
Model and suffix code 5
W
—
M55 MS CODE 6
Model and suffix code 6
W
—
M60 SELF CHECK
Self-diagnostics
R
See A60
—
P
Record
History of errors
P10 ERROR REC 1
Last error
W See A60
GOOD
—
P12 ERROR REC 2
Second recent error
W See A60
GOOD
—
P14 ERROR REC 3
Third recent error
W See A60
GOOD
—
P16 ERROR REC 4
Forth recent error
W See A60
GOOD
—
P60 SELF CHECK
Self-diagnostics
R
See A60
—
T
CHARACTERIZR Signal characterizer setting
T10 S. C. ENABLE
Signal characterizer permission W INHIBIT or ENABLE
INHIBIT
—
T11 NUM OF POINT
Number of coordinates
W 0 to 9
0
—
T20 X START (FIX)
Start point of X
R
0.00%
—
T21 Y START (FIX)
Start point of Y
R
0.00%
—
T22 X1
Coordinate 1 of X
W 0.00 to 100.00%
10.00
—
T23 Y1
Coordinate 1 of Y
W 0.00 to 100.00%
10.00
—
T24 X2
Coordinate 2 of X
W 0.00 to 100.00%
20.00
—
T25 Y2
Coordinate 2 of Y
W 0.00 to 100.00%
20.00
—
T26 X3
Coordinate 3 of X
W 0.00 to 100.00%
30.00
—
T27 Y3
Coordinate 3 of Y
W 0.00 to 100.00%
30.00
—
T28 X4
Coordinate 4 of X
W 0.00 to 100.00%
40.00
—
T29 Y4
Coordinate 4 of Y
W 0.00 to 100.00%
40.00
—
T30 X5
Coordinate 5 of X
W 0.00 to 100.00%
50.00
—
T31 Y5
Coordinate 5 of Y
W 0.00 to 100.00%
50.00
—
T32 X6
Coordinate 6 of X
W 0.00 to 100.00%
60.00
—
T33 Y6
Coordinate 6 of Y
W 0.00 to 100.00%
60.00
—
T34 X7
Coordinate 7 of X
W 0.00 to 100.00%
70.00
—
T35 Y7
Coordinate 7 of Y
W 0.00 to 100.00%
70.00
—
T36 X8
Coordinate 8 of X
W 0.00 to 100.00%
80.00
—
T37 Y8
Coordinate 8 of Y
W 0.00 to 100.00%
80.00
—
T38 X9
Coordinate 9 of X
W 0.00 to 100.00%
90.00
—
T39 Y9
Coordinate 9 of Y
W 0.00 to 100.00%
90.00
—
T40 X END (FIX)
End point of X
R
100.00%
—
T41 Y END (FIX)
End point of Y
R
100.00%
—
T60 SELF CHECK
Self-diagnostics
R
See A60
—
*1: R/W: R = Read only, W = Read & Write
*2:
The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.