Yokogawa EJA440E User Manual
Page 36
<5. Parameter Summary>
5-4
IM 01C25T03-01E
No. Parameter name
Item
*
1
R/W
Content
Default value
Applicable
model
U
pl
oa
d
da
ta
F
P
L
J
ADJUST
Adjusting data
J09 ADJ UNIT
Pressure adjusting unit select
W % or PRES UNIT
PRES UNIT
—
J10 ADJ PRES
Adjustment reference pressure
R
Unit specified in J09
—
J11 P ZERO ADJ
Automatic zero adjustment
W -32000 to 32000, unit
specified in J09
0.00000 kPa
—
J12 P SPAN ADJ
Automatic span adjustment
W -32000 to 32000, unit
specified in J09
100.000 kPa
—
J15 P ZERO DEV
Manual zero adjustment
W -32000 to 32000, unit
specified in J09
100.000 kPa
—
J16 P SPAN DEV
Manual span adjustment
W -32000 to 32000, unit
specified in J09
100.000 kPa
—
J20 ADJ SP
Adjustment reference pressure
of SP
R
Unit specified in J09
—
—
J21 SP ZERO ADJ
Automatic SP zero adjustment
W -32000 to 32000, unit
specified in J09
0.00000 MPa
—
—
J22 SP SPAN ADJ
Automatic SP span
adjustment*
2
W -32000 to 32000, unit
specified in J09
—
—
J25 SP ZERO DEV
Manual SP zero adjustment
W -32000 to 32000, unit
specified in J09
0.00000 MPa
—
—
J26 SP SPAN DEV
Manual SP span adjustment
W -32000 to 32000, unit
specified in J09
0.00000 MPa
—
—
J40 OUTPUT 4mA
4 mA adjustment
W -10.000 to 10.000%
0.000%
—
J41 OUTPUT 20mA
20 mA adjustment
W -10.000 to 10.000%
0.000%
—
J45 AMP TEMP
Amplifier temperature
R
Unit specified D40
0 degC
—
J50 ADJ WHO
Adjustment information
W 8 alphanumeric characters
—
J51 ADJ DATE
Adjustment information
W 16 alphanumeric characters
—
J52 ADJ LOC
Adjustment information
W 8 alphanumeric characters
—
J53 ADJ DESC
Adjustment information
W 16 alphanumeric characters
—
J55 EXT ZERO ADJ
External zeroing permission
W INHIBIT or ENABLE
ENABLE
—
J56 CLEAR ADJ
Clear adjustment
W ---, PRES, SP, 4-20mA, or
ALL
---
—
J60 SELF CHECK
Self-diagnostics
R
See A60
—
K
TEST
Test parameters
K10 OUTPUT X %
Test output % setting
W Within a range between D20
and D21
0.00%
—
K40 DO TEST
Test contact output
W OFF or ON
OFF
—
K45 TEST TIME
“OUTPUT X %” and “DO
TEST” duration time selection
W 10 min, 30 min, 60 min,
3 hour, 6 hour, 12 hour
10 min
—
K50 TEST KEY1
Special maintenance
parameter
W
—
K51 TEST KEY2
Special maintenance
parameter
W
—
K52 TEST KEY3
Special maintenance
parameter
W
—
K53 TEST KEY4
Special maintenance
parameter
W
—
K60 SELF CHECK
Self-diagnostics
R
See A60
—
M
DEVICE INFO
Device information
M10 SERIAL NO.
Serial number
R
—
M11 MFTR. DATE
Manufactured date
R
—
M12 EXTRA NO.
Customization number
R
—
M15 SOFT REV
Software revision
R
—
M16 BRAIN REV
BRAIN protocol revision
R
—
M17 MEMO1
Memo
W 16 alphanumeric characters
—
M18 MEMO2
Memo
W 16 alphanumeric characters
—
M19 MEMO3
Memo
W 16 alphanumeric characters
—
M20 ISOL MATL
Capsule material
W
—
M21 FILL FLUID
Fill fluid
W
—
*1: R/W: R = Read only, W = Read & Write
*2:
The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.