ZOLL M Series Defibrillator Rev YH User Manual
Page 86

M S
ERIES
O
PERATOR
’
S
G
UIDE
B-2
5. “POOR LEAD CONTACT” message.
•
Check that ECG cable is connected to patient and instrument.
•
Check that ECG electrodes are making good contact and not
dried out.
•
If changing from five (5) lead ECG patient cable to three (3)
lead ECG patient cable, turn unit OFF for at least 10 seconds.
•
Replace ECG cable.
6. Noisy ECG, Artifact, Wandering Baseline.
•
Consider 1 – 21Hz filter bandwidth (see M Series
Configuration Guide)
•
Prepare the patient’s skin prior to electrode attachment.
•
Check for proper adhesion of electrodes to patient.
•
Reduce or eliminate ECG artifact due to electrode or patient
cable movement. Route cables so that they don’t pull on
electrodes or swing excessively.
•
Ensure patient is motionless.
•
Check for possible excessive radio frequency interference.
7. Poor ECG signal level, calibration pulse normal.
•
Select another lead.
•
Apply new electrodes using different placement.
8. Inconsistent QRS beep or heart rate.
•
Increase beeper volume.
•
Select another lead.
•
Alter ECG electrode placement.
9. Sync marker is absent or inconsistent with QRS
waveform on display and recorder.
•
Ensure device is in SYNC mode.
•
Change ECG lead selection.
•
Alter ECG electrode placement.
•
Paper too narrow. It should be 90 mm wide.
10.
Unit displays the CABLE FAULT or PADDLE
FAULT message.
•
Check the connection between the Multifunction cable and
the MFE pads.
•
Remove the CPRD-to-MFC connector, if in use, and plug the
multifunction cable directly into the CPR-D•padz.
Symptom
Recommended Action