Electromagnetic immunity declaration (eid) – ZOLL M Series Defibrillator Rev YH User Manual
Page 80

M S
ERIES
O
PERATOR
’
S
G
UIDE
A-6
Electromagnetic Immunity Declaration (EID)
Guidance and manufacturer’s declaration — electromagnetic immunity for the M Series.
The M Series is intended for use in the electromagnetic environment specified below. The customer or user of the
M Series should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6kV contact
±8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
Not applicable
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 seconds
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 seconds
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the M Series
requires continued operation during
power mains interruptions, it is
recommended that the M Series be
powered by an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field.
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of typical
location in a typical commercial or
hospital environment.
NOTE: U
T
is the ac mains voltage prior to the application of the test level.