200j sel, Enter access code, 200j sel. confirm manual mode – ZOLL M Series Defibrillator Rev YH User Manual
Page 45

Automated External Defibrillator (AED) Operation
5-5
AED Manual Mode Operation
Press the Manual Mode softkey on the front panel of the
unit to enter the manual mode of operation.
Based on whether or not the device has been configured
with an access code, one of the following two screens
will be displayed.
In the above display, a three-digit access code must be
entered to enter the manual mode (if configured). Press
the individual softkey corresponding to the digit to be
entered (each digit entered must be between 0 and 3).
The highlight will automatically move to the next space.
Repeat until the access code is entered correctly and the
unit enters manual mode. If the Return to Auto softkey
is pressed, the unit will return to the semi-automatic
mode of operation.
If the access code is not configured, the message
“Confirm Manual Mode” and Confirm softkey will be
displayed. Press the Confirm softkey to enter the manual
mode of operation. If the Confirm softkey is not pressed
within 5 seconds the unit will revert back to semi-
automatic operation.
Refer to the “Manual Defibrillation”, “Advisory
Defibrillation”, “Pacing”, and “ECG Monitoring” sections
for the appropriate operation in manual mode.
AED Voice Prompts
The following is a list of the AED voice prompts:
•
ATTACH PADS
•
STAND CLEAR
•
PRESS SHOCK
•
CHECK PADS
•
CHECK PULSE
•
CHECK PATIENT
•
IF NO PULSE, PERFORM CPR
•
PRESS ANALYZE
•
NO SHOCK ADVISED
Troubleshooting
If your M Series unit does not function as expected, see
the Defibrillator Troubleshooting section starting on
page B-5.
200J SEL.
ID#
Manual
Mode
00:01
ECG
NIBP
Menu
0
ENTER ACCESS CODE
1
2
3
Return
to Auto
1
1
00:01
ECG
200J SEL.
CONFIRM MANUAL MODE
Confirm
ECG
00:01