ECG Monitoring
CPR Monitoring
Display
Patient Connection
3-lead ECG cable, 5-lead ECG cable, paddles or MFE Pads. Selectable by front panel switch.
Input Protection
Fully defibrillator protected. Special circuit prevents distortion of ECG by pacer pulse. (Pacer
version only).
Implanted
Pacemaker Spike
Display
Dedicated circuitry detects most implanted pacemaker spikes and provides standard display
marker of spike on ECG trace.
Bandwidth
0.5 Hz - 21 Hz (-3 dB) standard/0.05 Hz - 150 Hz Diagnostic
0.5 Hz - 27 Hz and 1 Hz - 21 Hz user-configurable
Lead Selection
Displayed on monitor
ECG Size
0.5, 1, 1.5, 2, 3 cm/mV - display on monitor
Heart Rate
Digital display 0 bpm - 300 bpm ±5%
Heart Rate Alarm
On/Off displayed on monitor. User-selectable, Tachycardia 60 bpm - 280 bpm,
Bradycardia 20 bpm - 100 bpm
1 Volt ECG Out
1.0 Volt/cm of deflection on stripchart recorder.
< 25 ms delay from patient ECG input.
Display Format
Non-fade moving bar display.
Compression Depth
0.75 to 3 inches ±0.25 inches
1.9 to 7.6 cm ±0.6 cm
Compression Rate
50 to 150 compressions per minute
Screen Type
High resolution, electoluminescent or liquid crystal display (LCD)
Screen Size
5.66 inches (14.4 cm) diagonally (EL Display)
Sweep Speed
25 mm/sec
Viewing Time
4 seconds
Messages
Erasing Report, Insert Card, Card Full, Replace Battery, Low Battery, Perform CPR,
ECG Too Large, Noisy ECG, Retry Analysis, Check Patient, Analysis Halted, Press Analyze,
No Shock Adv., Check Pulse, Shock Advised, Press Charge, Select Pads,
Select ECG Leads, Select Defib Mode, VF Alarms Off, Disable Sync, Analysis Restarted,
Check Pads, Poor Pad Contact, Defib Pad Short, Paddle Fault, ECG Lead Off,
Use Paddle DIschg, Open Air Discharge, Cannot Charge, Release Shock, 50 J Max,
Press Shock, Test OK, Test Fail, Pacer Disabled, Defib Disabled, Set Pace mA,
Set Pace Rate, Check Recorder, Analyzing ECG.