Operating controls and indicators, Pacer format (pacer version only), Heart rate alarm activated format – ZOLL M Series Defibrillator Rev YH User Manual
Page 27: Recorder on format

Operating Controls and Indicators
2-5
Pacer Format (Pacer version only)
Summary Report records 6 seconds of pre-pacer patient ECG data. Also recorded are the ECG lead, ECG size,
patient's heart rate, time and date.
After establishing a paced rhythm, turning the recorder on briefly will record the paced rhythm for later reports. If async
pace is active, the annotation “ASYNC PACE” is also recorded.
Heart Rate Alarm Activated Format
Summary Report records 6 seconds of pre-alarm patient ECG. Also recorded are the ECG lead, ECG size, patient’s
heart rate, time, and date. If the pacer is on during this event the pacing rate and pacing current are also recorded.
VF Alarm Activated (Refer to Section 8)
The summary report records 15 seconds of patient ECG data associated with each “VF” alarm. Also recorded are the
shock count, ECG lead, ECG size, patient’s heart rate, and noise events.
Recorder On Format
Summary Report records 6 seconds of patient ECG prior to turning on the recorder. Also recorded are the ECG lead,
ECG size, patient’s heart rate, time, and date. If the Pacer is on during this event the pacing rate and pacing current
are also recorded. If async pace is active, the annotation “ASYNC PACE” is recorded. AED units will additionally
include shock count and AED mode annotations.
PRE-PACE ECG
ECG LEAD
I I
NAME
ECG SIZE
2
15:19:50 10 MAY 93
HEART RATE
38
RECORDED ECG
ECG LEAD
I I
NAME
ECG SIZE
2
15:19:50 10 FEB 98
HEART RATE
38
PACE RATE ppm 62
PACE CURRENT mA 38
HEART RATE ALARM
ECG LEAD
LEAD I I
NAME
ECG SIZE
2
15:21:47 10 MAY 93
HEART RATE
41
AUTO DEFIB MODE
SHOCK COUNT
1
ECG LEAD
ELECTRODES
ECG SIZE
1
HEART RATE
283
NOISE EVENTS
0
CHECK PATIENT
SUMMARY COMPLETE
NAME
15:19:50 10 FEB 98
CHECK PATIENT
DEFIB ADVISORY
RECORDED ECG
ECG LEAD
I I
NAME
ECG SIZE
2
15:19:50 10 FEB 98
HEART RATE
45