Operator’s Shift Checklist for M Series Products (Manual)
Recommended checks and procedures to be performed at
the start of each shift. For more detailed information, see
the M Series Operator’s Guide.
Date
______________________________________________
Location ____________________________________________
Unit Serial Number ___________________________________
1. Condition
1st
Shift
2nd
Shift
3rd
Shift
Remarks
Unit clean, no spills, clear of objects on top, case intact
2. Multi-function Pads
1 set pre-connected / 1 spare
3. Paddles
Paddles clean, not pitted
Release from housing easily
4. Inspect cables for cracks, broken wires, connector
A ECG electrode cable, connector
B Defibrillator paddle cables
C Multi-function cable, connector
5. Batteries
A Fully charged battery in unit
B Fully charged spare battery available
6. Disposable supplies
A Electrode gel or gel patches
B MFE Pads in sealed pouches — 2 sets
C ECG electrodes
D Recorder paper
E Alcohol wipes
F Razors
7. Operational checks
A Power On Sequence
Turn unit to MONITOR, 4-beep tone heard
“MONITOR” message on display
ECG size X 1
“PADDLES” or ‘PADS” as lead selected
B Pacer Operation (Pacer Version ONLY)
Multi-function cable not connected to Test Connector
Turn to PACER, set pacer rate to 150 ppm, press RECORDER button
Pacer pulses occur ever 2 large divisions (10 small divisions)
Press 4:1 button, pulses occur every 8 large divisions
Set PACER OUTPUT to 0 mA, no “CHECK PADS” message
Set PACER OUTPUT to 16 mA, “CHECK PADS” message and alarm
Reconnect Multi-function cable to test connector.
Press Clear Pace Alarm softkey; “CHECK PADS” message disappears and
Pace alarm stops.
C Defibrillator
Multi-function cable connected to test connector: Set defib energy level to
30 joules, press SHOCK button; “TEST OK” message on Recorder
D Paddles
Paddles in holder: Set defib energy level to 30 joules, press paddles firmly into
the side wells, and simultaneously press and hold both defib discharge buttons;
“TEST OK” message on Recorder.
E Recorder
Press RECORDER button; Recorder runs. Press again; Recorder stops.
Inspect Recorder printing
8. Please check the appropriate box after each use of this checklist.
Signatures
No action required
1st ___________________
Minor problem(s) corrected
Disposable supplies replaced
2nd __________________
Major problem(s) identified — UNIT OUT OF SERVICE
3rd ___________________