ZOLL AED Plus Fully Automatic Rev B User Manual
Page 42
32
Fully Automatic AED Plus Administrator’s Guide
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
<5% Ut (>95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95%dip
in Ut) for 5 sec
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment
NOTE Ut is the a.c. mains voltage prior to application of the test level.
The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used
in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the Fully
Automatic AED Plus, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
a
3 Vrms
d = 1.17
10 Vrms
150 kHz to 80 MHz
in ISM bands
a
10 Vrms
d = 1.20
P
P