Power on self test, Manual self test – ZOLL AED Plus Fully Automatic Rev B User Manual
Page 27
![background image](/manuals/355830/27/background.png)
Fully Automatic AED Plus Administrator’s Guide
17
Power On Self Test
The Fully Automatic AED Plus performs a self test whenever the unit is turned on and verifies the
following functions:
1.
Battery Capacity: Verifies that the battery usage indicator shows adequate battery capacity
remaining.
2.
Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-
connected to the device.
3.
ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.
4.
Defibrillator Charge and Discharge Circuitry: Verifies that the device’s defibrillator electronics
are functional and can charge and discharge at 2 joules.
5.
Microprocessor Hardware/Software: Verifies proper function of the Fully Automatic AED Plus
microprocessor electronics and the integrity of its software.
6.
CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are
functional.
7.
Audio Circuitry: Verifies that voice prompts are functional.
Manual Self Test
You can initiate a manual self test on the Fully Automatic AED Plus by pressing and holding the unit’s
On/Off button for 5 seconds. The Fully Automatic AED Plus illuminates all graphic indicators and
issues voice and LCD display messages to allow user verification of the device’s visual and auditory
output functionality. In addition, the LCD shows the application software version currently running
on the unit.
This self test verifies the following Fully Automatic AED Plus functions:
1.
Battery Capacity: Verifies that the battery usage indicator shows adequate battery capacity
remaining.
2.
Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-
connected to the device.
3.
ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.
4.
Defibrillator Charge and Discharge Circuitry: Verifies that the device’s defibrillator electronics
are functional and can charge and discharge at 200 joules.
5.
Microprocessor Hardware/Software: Verifies proper function of the Fully Automatic AED Plus
microprocessor electronics and the integrity of its software.
6.
CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are
functional.
7.
Audio Circuitry: Verifies that voice prompts are functional.
8.
Display: Verifies that visual indicators are functional.
00:12
SHOCKS: 00
B:03.31 P:05.10
L:05.00 C:02.00
H:10.00
Elapsed Time
Shock Count
Language File Version
Boot File Version
Application Software Version
Firmware Version
Hardware Version