ZOLL AED Plus Fully Automatic Rev B User Manual
Page 17

Fully Automatic AED Plus Administrator’s Guide
7
ATTACH DEFIB PADS TO
PATIENT’S BARE CHEST.
Attach defibrillation pads to the victim’s bare chest.
CHECK ELECTRODE PADS.
Previously attached electrodes are not making good contact with
the victim’s skin or the electrodes are defective.
ADULT PADS.
Fully Automatic AED Plus has detected adult electrode pads
connected to it and adjusted defibrillation energy settings to adult
levels.
PEDIATRIC PADS.
Fully Automatic AED Plus has detected pediatric electrode pads
connected to it and adjusted defibrillation energy settings to
pediatric levels.
DON’T TOUCH PATIENT,
ANALYZING.
Do not touch victim, an ECG rhythm analysis is in progress or
about to begin.
SHOCK ADVISED.
ECG rhythm analysis has detected the presence of ventricular
fibrillation or shockable ventricular tachycardia.
NO SHOCK ADVISED.
ECG rhythm analysis has detected a rhythm that is not treatable
by defibrillation.
ANALYSIS HALTED. KEEP
PATIENT STILL.
ECG rhythm analysis has been halted due to excessive ECG
signal artifact. Stop any ongoing CPR and keep the victim as
motionless as possible.
SHOCK WILL BE DELIVERED
IN THREE (TWO), (ONE)
A shock is about to be delivered to the victim. Warn all persons
attending the victim to stand clear and stop touching the victim.
If the unit must be discharged before a shock is delivered, press
the On/Off button.
SHOCK DELIVERED.
A defibrillation shock has just been delivered to the victim.
NO SHOCK DELIVERED.
No shock was delivered to the victim because an error condition
was detected.
n SHOCKS DELIVERED.
A total of n shocks have been delivered since the Fully
Automatic AED Plus was turned on.
START CPR.
Begin CPR.
CONTINUE CPR.
Continue providing CPR. This prompt may also be issued if
Real CPR Help fails to detect chest compressions at least ¾ of an
inch deep.
PUSH HARDER.
CPR compressions are consistently less than 2 inches deep.
GOOD COMPRESSIONS.
After prompting to Push Harder, rescuer has succeeded in
delivering chest compressions at least 2 inches deep.
STOP CPR.
Stop CPR, the Fully Automatic AED Plus is about to begin an
ECG rhythm analysis.
Voice Prompt
Definition