ZOLL AED Plus Fully Automatic Rev B User Manual
Page 31

Fully Automatic AED Plus Administrator’s Guide
21
• displays a red “X” in the status indicator window, indicating that the batteries are at low capacity
or that the Fully Automatic AED Plus has failed other self tests.
Table 3: Battery Condition
Battery Condition
Indications
Correction
Low Battery with Fully
Automatic AED Plus off.
Audible beep from Fully
Automatic AED Plus once every
minute.
Replace batteries.
Low Battery during power
up self test.
CHANGE BATTERIES prompt
(when Fully Automatic AED
Plus is powered on)
Replace batteries.
Low Battery or other self
test failure with Fully
Automatic AED Plus
powered off or during self
test.
Status Indicator has red “X”
indicating failure to operate
(when off).
Replace batteries. Check or
replace electrodes. If red “X”
remains, return to ZOLL
Technical Support for service.
Low Battery with Fully
Automatic AED Plus
powered on.
CHANGE BATTERIES prompt
(unit powered on).
Replace batteries as soon as
possible.
Dead Battery
Status Indicator has red “X”
indicating failure to operate
(when off).
Replace batteries. If red “X”
remains, return to ZOLL
Technical Support for service.