16

Fully Automatic AED Plus Administrator’s Guide

Using the Self Test Feature

The Fully Automatic AED Plus performs the following self tests to verify unit integrity and its readiness

for emergency use:

• Battery Installation Self Test

• Power On Self Test

• Manual Self Test

• Automatic Self Test

• Automatic Monthly Test

Following successful completion of all self tests, the Fully Automatic AED Plus’s status indicator

displays a green check (

)

to show that all tests passed and that the unit is ready to use.

If the status indicator displays a red X following completion of any self test, the Fully Automatic AED

Plus is not ready for use and may be defective. Remove the Fully Automatic AED Plus from service

and consult the Troubleshooting section of this guide to help determine the problem.

Battery Installation Self Test

The Fully Automatic AED Plus performs a self test whenever batteries are installed and verifies the

following functions:

1.

Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-

connected to the device.

2.

ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.

3.

Defibrillator Charge and Discharge Circuitry: Verifies that the device’s defibrillator electronics are

functional and can charge and discharge at 200 joules.

4.

Microprocessor Hardware/Software: Verifies proper function of the Fully Automatic AED Plus

microprocessor electronics and the integrity of its software.

5.

CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are

functional.

6.

Audio Circuitry: Verifies that voice prompts are functional.

At the end of this self test, the Fully Automatic AED Plus prompts the user to press the Battery Reset

Button, located inside the battery compartment. Pressing this button resets the unit’s battery usage

indicator to full charge.

CAUTION! DO NOT press the Battery Reset Button unless all batteries are new. Pressing the Battery Reset

Button when used batteries are installed may result in a false high reading of battery capacity. See
the “Installing or Replacing Batteries” on page 18 for more information.