A.7 battery environmental – ZOLL AutoPulse Rev D Users Guide User Manual
Page 76

Page A-4
P/N 11440-001 Rev. 3
User Guide
A.7
Battery Environmental
Table A-5
Battery Specifications
Category
Specifications
Operating temperature
+32º to +113ºF (0º to +45ºC) ambient installed in device
Charge temperature
+41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)
Storage temperature
•
-4º to +77ºF (-20º to +25ºC) ambient for less than six months (may require
test-cycle to meet performance characteristics)
•
+77º to +95ºF (+25º to +35ºC) ambient for less than two months (may
require test-cycle to meet performance characteristics)
Operating altitude
0 to 15,000 ft. (0 to 4,572 m)
Enclosure protection
Meets IP24 per IEC 60529
Shock
Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration
Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s
2
)
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Free fall
Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.
Table A-6
Battery EMI/EMC Specifications
Title
Standard
Electrostatic discharge
IEC 61000-4-2, Level 3
Radiated emissions
CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A