beautypg.com

2 cleaning the autopulse platform, 3 storing the autopulse platform, Figure 4-2 battery charger control panel -2 – ZOLL AutoPulse Rev D Users Guide User Manual

Page 64

background image

Page 4-2

P/N 11440-001 Rev. 3

User Guide

Figure 4-2

Battery Charger Control Panel

4.

The Battery Charger will automatically perform a battery test-cycle when needed. To manually run
a test-cycle on a Battery, simply press the Battery Charger’s Start Test button at any time that the
Battery Charger is in Charge mode.

Note:

Do not remove a Battery from the Battery Charger until its charging completes or the
Battery’s runtime will be reduced.

Note:

Do not remove a Battery during a test-cycle or the Battery’s runtime will be unknown.
Removing a Battery during a test-cycle may cause the Battery Charger to automatically
enter a test-cycle mode the next time a Battery is inserted into it.

5.

Remove the Battery from the Battery Charger by gripping the Battery while pressing the finger latch
to disengage the locking bar. Pull the Battery straight out until it fully clears the charging bay.

Note:

Recently-charged Batteries can be warm to the touch. This is a consequence of normal
operation.

4.2

Cleaning the AutoPulse Platform

1.

Remove and dispose of the LifeBand.

2.

Wipe all the surfaces of the AutoPulse Platform free of foreign matter and spills with a disinfectant
or bactericidal wipe. Check the vents to ensure that they are free and clear of any obstructive matter.

Caution: Do not submerge the AutoPulse in liquid.

3.

Ensure that the AutoPulse is dry before storing.

4.3

Storing the AutoPulse Platform

1.

The AutoPulse should be powered down before it is stored.

Note:

The AutoPulse will automatically power down after 10 minutes of inactivity.

2.

The AutoPulse should be returned to a Carry Case, Hygiene Barrier or Transporter before being
stored. To purchase a Carry Case, Hygiene Barrier or Transporter, please contact your ZOLL
representative by calling +1.800.348.9011 or +1.978.421.9655.

3.

The AutoPulse should be stored in a cool, dry place.

4.

The Battery should be stored in the Battery Charger unless the Battery is deployed in an AutoPulse.