Page 3-6

P/N 11440-001 Rev. 3

User Guide

7.

Position the patient so that he/she is centered laterally (from left to right) and that the armpits are
aligned with the AutoPulse using the yellow line positioning guides on the platform as shown in
Figure 3-6.

Figure 3-6

Patient Alignment

Warning: Do not place or position the patient on the AutoPulse Platform in either a facedown

orientation or on the patient's side. Also ensure the head orientation is correct. The head
label arrow points towards the patient's head.

ON/OFF

Center the patient on the
AutoPulse Platform.

Line up armpits to
the yellow Patient
Alignment Reference.