C.2 nimh battery physical and environmental, Category, Specifications – ZOLL AutoPulse Rev A User Manual
Page 44

Page C-2
P/N. 12457-001 Rev. 2
Power System User Guide
C.2
NiMH Battery Physical and Environmental
Radiated emissions
Meets CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
Radiated Immunity
Meets IEC-61000-4-3, 80-2500 MHz, Level 3
Safety
Meets IEC-60601-1 including UL310DV.1.1 for Lithium batteries
Table C-2 NiMH Battery Specifications (Page 1 of 2)
Category
Specifications
Manufacturer
Manufactured for ZOLL Circulation
Model Number
8700-0702-01
Size (L
WH)
11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm).
Weight
5.1 lbs. (2.3 kg).
Type
Rechargeable Nickel-Metal Hydride (NiMH)
Battery voltage (nominal)
32.4V
Capacity
3200 mAh (typical)
Initial Battery run time
(nominal patient)
30 minutes (typical)
Maximum Battery charge
time
Less than 6¼ hours at 77ºF (25ºC)
Battery Test-Cycle time
Less than 12 hours per Test-Cycle session
Required replacement interval 100 full charge/discharge cycles.
Note:
The Battery will not operate after 100 full charge/discharge cycles.
Operating temperature
+32º to +113ºF (0º to +45ºC) ambient installed in device
Charge temperature
+41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)
Storage temperature
-4º to +95ºF (-20º to +35ºC) ambient for up to six months with charging every
four weeks, starting with a fully charged Battery.
Operating altitude
0 to 15,000 ft. (0 to 4,572 m)
Enclosure protection
Meets IP24 per IEC 60529
Shock
Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration
Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s
2
)
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Table C-1 Li-Ion Battery Specifications (Page 2 of 2)
Category
Specifications