Page 7-2

P/N. 12457-001 Rev. 2

Power System User Guide

3.

Check both of the fuses. If a fuse must be replaced, follow these steps:

a)

Replace both of the original fuses with T2.5A 250V AC fuses.

b)

Push in the fuse holder until the locking tab clicks into place.

c)

Re-attach the Power Cord.

d)

When the AutoPulse Multi-Chemistry Battery Charger is powered on, all the LED’s illuminate
briefly as the Charger performs a self test.

Note:

While the self test is being performed, if the indicator lights remain illuminated or if one
of the indicator lights fails to illuminate, contact ZOLL.