beautypg.com

Teledyne LeCroy QPHY-ENET User Manual

Page 78

background image

78

QPHY-ENET Operator’s Manual Rev G

Disturbing Signal Source

The signal generator used as the disturbing signal source must be set as follows. The exact amplitude of the
signal depends on the result from the calibration. See the

X

Disturber Calibration Procedure

X

(previous).

Mode 1 Disturbing Signal

CH1 : Waveform: SINE, Frequency: 31.25 MHz, Phase 0 degree

CH2 : Waveform: SINE, Frequency: 31.25 MHz, Phase 180 degree

Mode 4 Disturbing Signal

CH1 : Waveform: SINE, Frequency: 20.833 MHz, Phase 0 degree

CH2 : Waveform: SINE, Frequency: 20.833 MHz, Phase 180 degree

Recommended Generator

This type of test requires an Arbitrary Waveform Generator with dual-channel 250 MS/s waveform generation
capability. One such AWG is the TABOR Electronics Ltd. Model WW2572A. For more information on this AWG,
visit

HU

http://www.taborelec.com

U